The Boaray5000 series ventilator has two models: Boaray5000C and Boaray5000D, the differences of configuration are shown as below:
Configuration/Model |
Boaray5000C |
Boaray5000D |
VCV mode |
Yes |
Yes |
PCV mode |
Yes |
Yes |
PSV mode |
Yes |
Yes |
SIMV mode |
Yes |
Yes |
SPONT mode |
Yes |
Yes |
DualPAP mode(High/Low pressure) |
Optional |
Optional |
PRVC mode |
Optional |
Optional |
NPPV mode |
Optional |
Optional |
Tidal volume |
501500 mL |
202000 mL |
Maintenance schedule
Maintenance |
|
|
|
|
Maintenance |
Frequency |
|
|
Every patient |
|
Respiratory valve, reusable breathing tube |
Every day |
|
The ventilator surface |
Every half year |
|
Fan filter, air water filter, battery |
Every year |
|
Oxygen sensor |
When cleaning installation |
and |
Check the parts good or not, if necessary change or repair it |
General |
|
|
|
Contents |
|
information |
|
|
Classification |
b |
|
Electric shock defending type |
TYPE: , containing the internal power supply When you have doubt about the integrality of the external protective grounded or the protective grounded cable, the equipment should be replaced by the internal power supply (batteries) |
|
Electic shock |
|
|
|
TYPE B |
|
defending grade |
|
|
Suitabel range |
Pediatricweight: 5 20kg Adultweight: 20 150kg |
|
Power supply |
External power supply |
Power input: 100240Vfrequency 50/60Hz , Current input:1.5A (Max.) |
|
Backup batteries |
11.1VDC, 7800mAh |
|
Pressure supply range |
0.280.6MPa |
Gas supply |
|
The gas provided should not contain water, oil or impurity particle, |
|
Gas impurity |
and the content must be lower than the following standard. Air: H2O<7g/m3, Oil<0.5mg/m3 Oxygen: H2O<20mg/m3 |
Environment |
Working |
Temperature: +10 +50 Humdity: 95%non-condensation Atmospheric pressure: 50 kPa106 kPa |
|
|
|
|
|
Temperature: -20+55 |
|
Storage |
Humdity: 95%non-condensation |
|
|
Atmospheric pressure: 50 kPa106 kPa |
The whole device |
Trolley |
Volume: 707512956mm, Weight: 18.7kgnet weight |
|
|
|
|
Host |
Volume: 489443541mm, Weight: 23.2kgnet weight |
Technical parameters
Description Parameter
Pneumatic electronically contronlled ventilator Category
Display |
15inch, TFT display, touch screen (optional) |
|
Battery |
Li batteries, power supply for about 120 minutes. |
|
Respiration mode |
VCV, PCV, SIMV+VCV, SIMV+PCV, SIMV+PRVC, PSV, NPPV(optional), PRVC(optional), DualPAP(optional) |
SPONT, |
Graphics display |
Waveform: prssure-time, flow-time,volume-time. Loop graph(optional):pressure-volume,flow-volume,flow-pressure. Others(optional): CO2, SPO2 |
|
Safe pressure in the airway system |
125cmH2O. |
|
Data communication interface |
RS232 serial port, communication interface, LVDS interface |
|
Sound pressure levels(Normal operation) |
65dB (A) |
|
Setting
parameters
Parameter |
Description |
|
Accurancy |
Tidal volume |
Range: 202000 mL (D series), Range: 501500 mL (C series) Increment: 20100 Ml: 5 mL 1001000 mL: 10 mL 10002000mL: 50 mL |
|
20mL, or 15% of the set value, whichever is greater |
Respiration frequency |
Range: in SIMV mode: 140 bpm In other modes: 4100 bpm, increment: 1 bpm |
|
2bpm or 10% of the set value, whichever is greater |
Inspiration time |
Range: 0.110.0 s Increment: 0.1 s |
|
0.1s or 10% of the set value, whichever is greater |
Breathing hold time |
Range: 04 s, increment: 0.1 s |
|
|
Pressure trigger |
Range: -20cmH2O-1 Increment: 1 cmH2O |
|
<-5 cmH2O: 1 cmH2O, the other is 20% |
Flow trigger |
Range: 0.520 L/min Increment: 0.5 L/min |
|
<-5 L/min : 1 L/min, the other is 20% |
PEEP |
Range: 050 cmH2O Increment: 1 cmH2O |
|
2cmH2O or 10% of |
Pressure support |
Range: 170 cmH2O Increment: 1 cmH2O Ppeak (pressure support) = pressure (Pressure support) + PEEP |
setting |
|
Range: 570 cmH2O |
the set value, |
Pressure control |
Increment: 1 cmH2O Ppeak (pressure control) = pressure setting |
whichever is greater |
|
(pressure control) + PEEP |
|
High pressure level(optional) |
Range: 570 cmH2O Increment: 1 cmH2O Not suitable for C series, and optional for D series |
|
Low pressure level(optional) |
Range: 050 cmH2O Increment: 1 cmH2O Not suitable for C series, and optional for D series |
|
High pressure support(optional) |
Range: 170 cmH2O Increment: 1 cmH2O |
|
Low pressure support(optional) |
Range: 170 cmH2O Increment: 1 cmH2O |
|
Oxygen concentration |
Range: 21100% Increment: 1% |
6% of the set value |
Apnea time |
Range: 1060 sec Increment: 5 sec |
1 s or 10% of the set value |
Respiration sensitivity |
Range: 1060% Increment: 10% |
|
Prate |
Range: 15 |
|
|
Increment: 1 |
|
Monitoring
parameters
Parameter |
Description |
Inspiration tidal volume |
Range: 03000 mL, Resolution: 1 mL |
Expiration tidal volume |
Range: 03000 mL, Resolution: 1 mL |
Minute ventilation volume |
Range: 099 L, Resolution: 0.1 L |
Spontaneous minute ventilation |
Range: 099 L, Resolution: 0.1 L |
volume |
|
Total respiration frequency |
Range: 0100bpm, Resolution: 1 bpm |
Spontaneous respiration requency |
Range: 099 bpm, Resolution: 1 bpm |
Inspiration and expiration rate |
Range: 4:11:9, Resolution: 0.1 |
PEEP |
Range: 050 cmH2O, Resolution: 1 cmH2O |
Airway peek pressure |
Range: 0100 cmH2O, Resolution: 1 cmH2O |
Average airway pressure |
Range: 0100 cmH2O, Resolution: 1 cmH2O |
Inspiration platform pressure |
Range: 0100 cmH2O, Resolution: 1 cmH2O |
Minimum airway pressure |
Range: 0100 cmH2O, Resolution: 1 cmH2O |
Oxygen concentration |
Range: 21100%, Resolution: 1% |
Airway resistance |
Range: 0200 cmH2O/(L/s), Resolution: 1 cmH2O/(L/s) |
Compliance |
Range: 0200 mL/cmH2O, Resolution: 1 mL/cmH2O |
Rapid shallow breathing |
Range: 0.0-1000.0 bpm/L, Resolution: 1 bpm/L |
Occlusion pressure P0.1 |
Range: 30>=P0.1>= 0, Resolution: 0.1 cmH2O |
Negative inspiration force |
Range: -40~0cmH2O, Resolution: 0.1 cmH2O |
AutoPEEP |
Range: 0~100cmH2O, Resolution: 0.1 cmH2O |
PR |
Range: 30250 bpm, Resolution: 1bpm |
EtCO2 |
Range: 099 mmHg, Resolution: 1mmHg |
InCO2 |
Range: 099 mmHg, Resolution: 1mmHg |
Main alarm
parameters
Alarm content |
|
|
Range |
Tidal volume low |
High |
1002000 mL, OFF |
|
Low |
OFF, 202000 mL |
|
|
Minute volume |
High |
040 L |
|
Low |
OFF, 039 L |
|
|
O2 concentration |
High |
21100% |
|
Low |
OFF, 2199% |
|
|
Airway pressure |
High |
6100 cmH2O |
|
Low |
099cmH2O |
|
|
Frequency |
High |
1100bpm |
|
Low |
099 bpm |
|
|
EtCO2 (optional) |
High |
199 mmHg |
|
Low |
OFF, 098 mmHg |
|
|
SPO2 (optional) |
Low |
80100% |
|
O2 supply pressure |
High |
110750 KPa |
|
Low |
100740 KPa |
|
|
Air supply pressure |
High |
110750 KPa |
|
|
Low |
100740 KPa |
|
PRVC airway pressure upper limits |
PlimitPpeak+5cmH2O |
|
|
High Continues Pressure |
It will give an alarm when the airway pressure is PEEP+15cmH2O for 15 seconds |
continuous above |
|
Apnea |
Range:1060 s, increment :5 s |
|
|
O2 supply pressure low |
The O2 supply pressue is less than 0.28 MPa |
|
|
Air supply pressure low |
The O2 supply pressue is less than 0.28 MPa |
|
|
AC failure |
The alarm time is more than 120s |
|
|
Low battery capacity |
The battery working time is 10 minutes after alarm |
|
|
Battery exhausted |
The battery working time is 5 minutes after alarm |
|
|
Alarm silence count down |
100 s |
|
|
Alarm sound |
>65dB |
|
CO2module specificationoptional
Parameter |
Specification |
|
Measuring mode |
Mainstream |
|
Operation temperature |
0-40/ 32-104 H |
|
Storage and transportation temperature |
-40-75/ -40-167 H |
|
Operating humidity |
< 40 hPa H2O (non-condensing) (95 %RH at 30 C) |
|
Storage and transportation humidity |
5-100 % RH, condensing |
|
Operation atmospheric pressure |
525-1200 hPa (525hPa corresponding to an altitude of 4572 m/15000 feet) |
|
Storage and transportation pressure |
500 to1200 hPa |
|
Total system response time |
1 second |
|
Accuracy specifications(during |
Range |
Accuracy |
standard conditions |
0 to 114 mmHg |
(0.2 mmHg + 2% of reading) |